Electrical exposure risk associated with hands-on defibrillation

RESUSCITATION JOURNAL COVER

Electrical exposure risk associated with hands-on defibrillation

Daniel L. Lemkin, Michael D. Witting, Michael G. Allison, Ali Farzad, Michael C. Bond, Mark A. Lemkin

 

 

Abstract

 

Background

 

The use of hands-on defibrillation (HOD) to reduce interruption of chest compression after cardiac arrest has been suggested as a means of improving resuscitation outcomes. The potential dangers of this strategy in regard to exposing rescuers to electrical energy are still being debated. This study seeks to determine the plausible worst-case energy-transfer scenario that rescuers might encounter while performing routine resuscitative measures.

 

Methods

 

Six cadavers were acquired and prepared for defibrillation. A custom instrumentation-amplifier circuit was built to measure differential voltages at various points on the bodies. Several skin preparations were used to determine the effects of contact resistance on our voltage measurements. Resistance and exposure voltage data were acquired for a representative number of anatomic landmarks and were used to map rescuers’ voltage exposure. A formula for rescuer-received dose (RRD) was derived to represent the proportion of energy the rescuer could receive from a shock delivered to a patient. We used cadaver measurements to estimate a range of RRD.

 

Results

 

Defibrillation resulted in rescuer exposure voltages ranging from 827?V to ?200?V, depending on cadaver and anatomic location. The RRD under the test scenarios ranged from 1 to 8?J, which is in excess of accepted energy exposure levels.

 

Conclusions

 

HOD using currently available personal protective equipment and resuscitative procedures poses a risk to rescuers. The process should be considered potentially dangerous until equipment and techniques that will protect rescuers are developed.